应用定量CT测量肝脏脂肪含量的可重复性分析

Reproducibility of liver fat content measured using quantitative computed tomography

摘要:

目的:对在体检人群中应用定量CT(QCT)技术测量肝脏脂肪含量的可重复性进行评估,为QCT应用于临床肝脏脂肪含量测量提供依据。方法:2019年4—7月,连续抽取在河南省人民医院健康管理科行QCT检查的291例体检者,男性214例(73.5%),女性77例(26.5%),年龄(48.7±11.0)岁。采用QCT工作站测量体检者的肝脏脂肪含量,由3位测量者(A、B、C)在QCT工作站上独立测量肝脏脂肪含量,测量者A在2周后重复测量1次,测量者A两次测量结果取均值作为其最终测量结果。采用均数±标准差描述计量资料,采用组内相关系数( ICC)评价测量者间和测量者内测量结果的可重复性。 结果:测量者A的第1次测量结果为(10.46±5.55)%,第2次为(10.66±5.59)%,最终结果为(10.56±5.51)%;测量者B、C测量结果分别为(10.70±5.45)%、(10.86±5.77)%。测量者A、B、C之间肝脏脂肪测量结果的 ICC=0.960(95% CI:0.951~0.967, P<0.001)。测量者A的2次肝脏脂肪测量结果 ICC=0.953(95% CI:0.941~0.962, P<0.001)。均 ICC>0.75,一致性均较好。 结论:在测量方法一致的前提下,采用QCT工作站测得的肝脏脂肪含量结果在测量者间和测量者内都具有良好的重复性,有一定的临床应用前景。

更多
abstracts:

Objective:To provide support for the clinical application of quantitative computed tomography (QCT) in the measurement of liver fat content, this study evaluated the intra-observer and inter-observer reproducibility of liver fat content measured by QCT in a population receiving physical examinations.Methods:From April to July 2019, 291 people were consecutively selected who underwent QCT examination in the health management department of Henan Provincial People′s Hospital. There were 214 males (73.5%) and 77 females (26.5%), aged 48.7±11.0. We measured liver fat content by QCT workstation. Three observers (A, B, C) measured their liver fat content independently, then observer A performed re-testing two weeks later. The mean value of the two measurements from observer A was taken as the final result. Measurement data were described by mean±SD. Intra-observer and inter-observer reproducibility were assessed using intra-class correlation coefficients ( ICC). Results:The first measurement result for observer A was 10.46±5.55 and the second measurement for observer A was 10.66±5.59, resulting in a final value of 10.56±5.51. The measurement results of observers B and C were 10.70±5.45 and 10.86±5.77, respectively. The ICC value of liver fat content values measured by the three observers was 0.960 (95% CI: 0.951-0.967, P<0.001) and the ICC value of liver fat content values for the two measurements of observer A was 0.953 (95% CI: 0.941-0.962, P<0.001). The ICC values were>0.75, so reproducibility of results was good. Conclusions:If the measurement method is consistent, the results for liver fat content measured by a conventional CT scanner and QCT workstation will have good reproducibility between and within observers, and will also have certain clinical application prospects.

More
作者: 陈白如 [1] 郭鸣周 [2] 屈敏 [3] 张潇琳 [3] 马会民 [3] 赵小娟 [3] 邹智 [4] 李永丽 [3]
期刊: 《中华健康管理学杂志》2020年14卷4期 318-321页 ISTICCSCD
栏目名称: 脂肪肝精准测评与管理
DOI: 10.3760/cma.j.cn115624-20191102-00477
发布时间: 2024-03-05
基金项目:
国家重点研发计划资助 河南省科技厅科技攻关项目 科技厅省科学自然基金 National Key Research and Development Program of China Technological Project of Henan Province Natural Science Foundation of Henan Province
  • 浏览:317
  • 下载:262

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷