Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements

摘要Background:Measuring total serum calcium is important for the diagnosis of diseases.Currently,results from commercial kits for calcium measurement are variable.Generally,the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes.However,the commutability of the EQA samples and calibrators is often unknown,which limits the effectiveness of EQA schemes.The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials.Methods:Inductively coupled plasma mass spectrometry was applied as a comparative method,and 14 routine methods were chosen as test methods.Forty-eight serum samples from individual patients and 25 processed materials were quantified.A scatter plot was generated from patient samples,and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays.Results:All assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99).For all assays,the mean bias for the 48 patient samples ranged from-0.13 mmol/L to 0.00 mmol/L (-5.61-0.01%),and the ranges for the three concentrations were-0.10-0.04 mmol/L (-5.71-2.35%),-0.14--0.01 mmol/L (-5.80--0.30%),and-0.19-0.04 mmol/L (-6.24-1.22%).The EQA samples,calibrators,and animal sera exhibited matrix effects in some assays;human serum pools were commutable in all assays;certificate reference materials were commutable in most assays,and only GBW09152 exhibited a matrix effect in one assay;and aqueous reference materials exhibited matrix effects in most assays.Conclusions:Biases for most assays were within the acceptable range,although the accuracy of some assays needs improvement.Human serum pools prepared from patient samples were commutable,and the other tested materials exhibited a matrix effect.

更多
广告
作者单位 National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine,Beijing 100730, China [1] Department of Clinical Laboratory, Children's Hospital, Capital Institute of Pediatrics, Beijing 100020, China [2] National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine,Beijing 100730, China;Graduate School, Peking Union Medical College, Chinese Academy of Medical Sciences, Beijing 100730, China [3]
栏目名称
DOI 10.4103/0366-6999.235109
发布时间 2018-08-13
提交
  • 浏览19
  • 下载4
中华医学杂志(英文版)

中华医学杂志(英文版)

2018年131卷13期

1584-1590页

SCIMEDLINEISTICCSCDCABP

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷