CBCT 引导前列腺癌 IMRT 图像配准及靶区外放边界分析

Image registration and target volume margins in cone-beam computed tomography-guided intensity-modulated radiotherapy for prostate cancer

摘要目的:通过分析千伏级 CBCT 引导前列腺癌 IMRT 的数据,为选择合理的图像配准方法和适宜的外放边界提供临床依据。方法针对16例接受根治性 IMRT 的前列腺癌患者,共行CBCT 在线校正治疗体位214次。采用常规皮肤标记激光对位后采集图像,将所获得 CBCT 图像与计划 CT 图像进行默认自动配准、骨性配准、软组织配准及手动靶区配准。比较4种配准方式之间差异,并计算由 CTV 外放产生 PTV 间距。结果16例患者默认自动配准、骨性配准、软组织配准及手动靶区配准方式在左右、前后、上下方向平移摆位误差分别为(-0.6±2.8)、(-0.6±4.5)、(-0.6±3.8) mm,(-0.7±2.7)、(-0.9±4.5)、(-0.8±4.1) mm,(-0.8±2.6)、(-0.3±4.4)、(-1.1±4.0) mm,(-0.6±2.9)、(-0.7±5.1)、(-0.9±3.9) mm。经分析4种配准方式之间相近。 PTV 在左右、前后、上下方向外放间距分别为4.7、5.2、6.5 mm。结论综合各种因素考虑,应用在线默认自动配准+手动微调CBCT 引导放疗技术治疗前列腺癌患者更为合适。 PTV 安全外放边界在左右、前后、上下方向分别为4.7、5.2、6.5 mm。

更多

abstractsObjective To analyze the data from intensity-modulated radiotherapy ( IMRT) for prostate cancer guided by kilovoltage cone-beam computed tomography (CBCT), and to provide a clinical basis for selecting the optimal image registration method and reasonable target volume margins.Methods A total of 16 patients with prostate cancer who received radical IMRT were enrolled, and CBCT for online position verification was performed 214 times.The images were obtained after conventional skin marking and laser alignment, and automatic registration, bone registration, soft tissue registration, and manual registration were performed for CBCT images and planned CT images.The differences between these four registration methods were evaluated, and the margins for extending clinical target volume into planning target volume (PTV) were calculated.Results The setup errors in left-right, anterior-posterior, and cranial-caudal directions for automatic registration, bone registration, soft tissue registration, and manual registration were-0.6±2.8 mm/-0.6±4.5 mm/-0.6±3.8 mm,-0.7±2.7 mm/-0.9±4.5 mm/-0.8±4.1 mm,-0.8± 2.6 mm/-0.3±4.4 mm/-1.1±4.0 mm, and-0.6±2.9 mm/-0.7±5.1 mm/-0.9±3.9 mm, respectively. There were no significant differences between the four registration methods.The margins for extension in the left-right, anterior-posterior, and cranial-caudal directions were calculated as 4.7 mm, 5.2 mm, and 6.5 mm, respectively.Conclusions With a comprehensive consideration of various factors, a default setting of automatic registration and manual fine adjustment is appropriate for CBCT-guided radiotherapy for prostate cancer.The margins for extension in the left-right, anterior-posterior, and cranial-caudal directions are 4.7 mm, 5.2 mm, and 6.5 mm, respectively.

More
广告
  • 浏览569
  • 下载229
中华放射肿瘤学杂志

中华放射肿瘤学杂志

2016年3期

249-254页

ISTICPKUCSCD

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷