二维半导体探测器阵列的方向性响应和在调强放射治疗合成剂量验证中的应用研究

The angular dependence of a 2-dimensional diode array and the feasibility of its application in verifying the composite dose distribution of intensity-modulated radiation therapy

摘要Background and Objective: The planning dose distribution of intensity-modulated radiation therapy (IMRT) has to be verified before clinical implementation. The commonly used verification method is to measure the beam fluency at 0 degree (0°) gantry angle with a 2-dimensional (2D) detector array, but not the composite dose distribution of the real delivery in the planned gantry angles. This study was to investigate the angular dependence of a 2D diode array (2D array) and the feasibility of using it to verify the composite dose distribution of IMRT. Methods: Angular response of the central detector in the 2D array was measured for 6 MV X-ray, 10 cm × 10 cm field and 100 cm source axis distance (SAD) in different depths. With the beam incidence angle of 0°-60°, at intervals of 10°, and inherent buildup of the 2D array (2 g/cm2), the array was Irradiated and the readings of the central diode were compared with the measurement of thimble ionization chamber. Using a combined 30cm×30cm×30cm phantom which consisted of solid water slabs on top and underlying the 2D array, with the diode detectors placed at 8 g/cm2 depth, measurements were taken for beam angles of 0°-180° at intervals of 10° and compared with the calculation of treatment planning system (TPS) that pre-verified with ion chamber measuring. Results: Differences between the array detector and thimble chamber measurements were greater than 1% and 3.5% when the beam angle was larger than 30° and 60°, respectively. The measurements in the combined phantom were different from the calculation as high as 20% for 90° beam angle, 2% at 90°± 5° and less than 1% for all the other beam angles. Conclusions: The 2D diode array is capable of being used in composite dose verification of IMRT when the beam angles of 90°±5° and 270°± 5° are avoided.

更多
广告
作者 李齐林 邓小武 陈立新 Xiao-Yan Huang Shao-Min Huang 学术成果认领
分类号 R73
栏目名称 Method and technology
发布时间 2010-07-02
基金项目
Wu Jie-ping Medical Foundation
  • 浏览137
  • 下载42
癌症

癌症

2010年29卷6期

617-620页

MEDLINEISTICCA

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷