Icare回弹式眼压计与Goldmann压平眼压计眼压测量结果的比较

Comparison of intraocular pressure measured by Icare rebound tonometer and Goldmann applanation tonometer

摘要背景 Icare回弹式眼压计作为一种新式眼压计,有必要对它的临床应用价值进行评估.目的 通过比较分析Icare回弹式眼压计和Goldmann压平眼压计(GAT)的眼压测量结果,探讨Icare的临床价值.方法可疑青光眼、青光眼、屈光不正及部分健康体检者78例共152眼同时接受Icare、GAT眼压测量,受检眼先行Icare测量,然后再进行GAT测量,2次测量间隔3~5 min.对比分析两种眼压计的测量结果.结果使用Ieare和GAT测得的眼压均值分别为(19.16±5.03)mmHg和(18.41±4.52)mmHg,96眼(63.2%)两者的眼压差值≤1 mmHg,二者的测量值差异虽有统计学意义,但二者的变化呈明显正相关(r=0.940,P<0.01).当Icare眼压测量值<16 mmHg时,Icare的眼压测量值低于GAT,而当Icare眼压测量值≥16 mmHg时恰好相反;CCT偏薄、正常以及偏厚的情况下,Icare的眼压测量值均高于GAT的眼压测量值.Icare、GAT的眼压测量值和CCT间呈正相关(r=0.341,P<0.01;r=0.333,P<0.01).结论 与GAT眼压计比较,Icare回弹式眼压计易操作,测量结果可靠.临床实用性更强.

更多

abstractsBackground As a new tonometer,it is necessary to assess the clinical value of Icare rebound tonometer.Objective This study was to compare the intraocular pressure(IOP)values measured by Icare with that measured by GAT,and discuss the clinical value of leare rebound tonometer. Methods IOP measurement was performed on 152 eyes of 78 subjects with suspicious glaucoma,glaucoma,refractive error and normal examinnee by Icare and GAT respectively.The Icare IOP was measured firstly and then the GAT IOP was carried out with the 3-or 5-minute interval.The IOP values were compared between Ieare and GAT.This study was approved by Ethic Committee of Wuhan General Hospital of Chinese PLA.Written informed consent was obtained from each subject prior to this study. Results The mean IOP values of Icare and GAT were(19.16±5.03)mmHg and(18.41±4.52)mmHg respectively.The differences between Icare IOP and CAT IOP were less than or equal to 1 mmHg in 96 of 105 eyes(63.2%).The positive correlation was found between the Icare IOP and GAT IOP(r=0.940,P<0.01).The Ieare IOP was lower than that of GAT when IOPIcare<16 mmHg,however,the IOP of Icare were higher when IOPIcare≥6 mmHg;the IOP of Icare were higher than that of GAT in the total CCT range.The correlation coefficients of IOP of Icare or CAT with CCT were 0.341(P<0.01)and 0.333(P<0.01),respectively. Conclusion Compared with GAT,Icare is more feasible in clinic because it is practicable and reliable.

More
广告
栏目名称
DOI 10.3760/cma.j.issn.2095-0160.2011.05.016
发布时间 2011-07-29(万方平台首次上网日期,不代表论文的发表时间)
  • 浏览450
  • 下载230
中华实验眼科杂志

中华实验眼科杂志

2011年29卷5期

456-459页

ISTICPKUCSCDCA

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷