Orbscan-Ⅱ、Pentacam及超声测厚仪角膜厚度测量值的比较

Comparison of central corneal thickness measured with Orbscan-Ⅱ, Pentacam and ultrasonic pachymetry

摘要目的 探讨Orbscan-Ⅱ、Pentacam及超声角膜测厚仪三种仪器测量角膜中央厚度的差异及其临床意义.方法 对预行准分子激光角膜手术的近视患者159例(318眼),采用Orbscan-Ⅱ、Pentacam及超声角膜测厚仪三种仪器测量角膜中央厚度,并对测量结果进行统计学分析.再根据超声角膜测厚仪所测得的角膜厚度进一步将患者分为三组(<500 μm组、500~570 μm组和≥570μm组),在每个组中对三种方法所测得的结果进行统计学分析.结果 Orbscan-Ⅱ、Pentacam、超声角膜测厚仪测得的角膜中央厚度平均值分别为(527.9±43.0)μm、(526.3±38.4)μm、(522.6±37.8)μm.经方差分析,三者差异无统计学意义(F=1.491,P=0.226).分组后,角膜厚度<500μm和500~570 μm组,三种方法所得数据差异无统计学意义(F=1.546,P=0.215;F=-2.107,P=0.123);而≥570μm组,三种方法测量结果差异存在统计学意义(F=5.396,P=0.006),进一步采用Bonferroni检验分析组间差异,Orbcan-Ⅱ组测量值明显高于Pentacam组(P=0.029)和超声测厚仪组(P=0.010),Pentacam组和超声测厚仪组之间无差异(P>0.05).结论 三种方法用于测量准分子激光角膜手术前近视患者的角膜中央厚度时虽然存在一定差异,但总体上无统计学意义,具有很好的协同性,但相互间还不能完全替代.

更多

abstractsObjective To assess the reliability of Orbscan-Ⅱ , Pentacam and ultrasonic pachymetry (US) for central corneal thickness (CCT) measurements. Methods Central corneal thick- ness was measured using Orbscan-Ⅱ, Pentacam and ultrasonic pachymetry in 159 preoperative cases (318 eyes). Patients were further divided into three groups according to corneal thickness (<500 μm group, 500~570 μm group and ≥570 μm group) as determined by ultrasonic pachymetry and the differences in the measurements obtained by all three methods were statistically analyzed for the three groups. Results The mean CCT readings with Orbscan-Ⅱ, Pentacam, and ultrasonic pachymetry were (527.9±43.0)μm, (526.3±38.4)μm, and (522.6±37.8)μm, respectively. No statistically significance differences were found among the three instruments (F=1.491, P=0.226) using analysis of variance. After the patients were divided into three groups based on the ultrasonic pachymetry measurements, the results showed that there was no statistically significant difference in measurements by the three instruments for the <500 μm group and the 500~570 μm group (F=1.546, P=0.215;F=2.107, P= 0.123), while there was a significant difference for the ≥570 μm group (F=5.396, P=0.006). Differ-ences among the groups using Bonferroni tests show that central corneal thickness (CCT) measured with Orbcan-Ⅱ was significantly thicker than that obtained with the Pentacam (P=0.029) and ultra-sonic pachymetry (P=0.010). No statistically significant difference was found between measurements with Pentacam and ultrasonic pachymetry. Conclusion There are some differences in the three meth-ods used to measure the CCT of patients before excimer laser myopic surgery, but overall there was no statistically significant difference, and the methods could be concordant in clinics. However, the three instruments still cannot completely replace one other.

More
广告
分类号 R77(眼科学)
栏目名称
DOI 10.3760/cma.j.issn.1674-845X.2010.01.018
发布时间 2010-06-07(万方平台首次上网日期,不代表论文的发表时间)
  • 浏览134
  • 下载218
中华眼视光学与视觉科学杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷