Signal interference between drugs and metabolites in LC-ESI-MS quantitative analysis and its evaluation strategy
摘要Liquid chromatography-electrospray ionization tandem mass spectrometry(LC-ESI-MS)is a widely utilized technique for in vivo pharmaceutical analysis.Ionization interference within electrospray ion source,occurring between drugs and metabolites,can lead to signal variations,potentially compromising quantitative accuracy.Currently,method validation often overlooks this type of signal interference,which may result in systematic errors in quantitative results without matrix-matched calibration.In this study,we conducted an investigation using ten different groups of drugs and their corresponding me-tabolites across three LC-ESI-MS systems to assess the prevalence of signal interference.Such in-terferences can potentially cause or enhance nonlinearity in the calibration curves of drugs and metabolites,thereby altering the relationship between analyte response and concentration for quanti-fication.Finally,we established an evaluation scheme through a step-by-step dilution assay and employed three resolution methods:chromatographic separation,dilution,and stable labeled isotope internal standards correction.The above strategies were integrated into the method establishment process to improve quantitative accuracy.
更多相关知识
- 浏览4
- 被引0
- 下载0

相似文献
- 中文期刊
- 外文期刊
- 学位论文
- 会议论文


换一批