三种仪器测量水平角膜直径的差异及其与睫状沟直径的一致性研究

Difference of horizontal corneal diameter measured by three devices and consistency horizontal sulcus to sulcus diameter

摘要目的:探讨三种常用仪器测量水平角膜直径的差异及其水平睫状沟直径的一致性。方法:横断面研究。选取苏州大学附属理想眼科医院于2018年1月至2021年12月行术前检查的屈光不正者870人(1 740眼)。年龄17~49(27.76±6.60)岁。所有受试者均行Pentacam、IOLMaster 700、OPD-Scan Ⅲ眼部生物测量,眼状沟直径应用UBM检查。分析前3种仪器所测量水平角膜直径之间的差异及与睫状沟直径之间的差异。结果:3种仪器所测水平角膜直径最大的是IOLMaster[(12.00±0.39)mm];其次是OPD-Scan Ⅲ[(11.85±0.40)mm];最小的是Pentacam[(11.59±0.38)mm],总体差异有统计学意义( F=281.69, P<0.001)。IOLMater所测得的水平角膜直径比水平睫状沟直径大(0.49±0.32)mm(差值的95%一致性界限:-0.14,1.13);OPD-Scan Ⅲ所测水平角膜直径与水平睫状沟直径的差值为(0.34±0.32)mm(95%LoA:-0.29,0.98);Pentacam所测水平角膜直径与水平睫状沟睫状沟直径的差值为(0.08±0.29)mm(95%LoA:-0.49,0.66)。所有仪器所测水平角膜直径与水平睫状沟的差值均与平均角膜曲率呈正相关,与眼轴呈负相关(均 P<0.01)。 结论:3种仪器测量角膜水平角膜直径与水平睫状沟直径均有差异,其中Pentacam最接近,而差异与平均角膜曲率和眼轴具有相关性。

更多

abstractsObjective:To investigate the difference of horizontal corneal diameter measured by three common devices.Methods:This was a cross-sectional study. A total of 1 740 eyes of 870 patients from Jan. 2018 to Dec. 2021 in Lixiang Eye Hospital of Soochow University were selected. The age ranged from 17 to 49 years, with an average of (27.76±6.60) years. All subjects were examined with Pentacam, IOLMaster 700, OPD-Scan Ⅲ and the horizontal sulcus to sulcus diameter were examined with UBM. The differences between the horizontal corneal diameters measured with the three instruments and the ciliary sulcus diameters were analyzed.Results:The largest horizontal corneal diameter was measured by IOLMaster, with an average of (12.00±0.39) mm; followed by OPD-Scan Ⅲ, with an average of (11.85±0.40) mm; and the smallest was Pentacam, with an average of (11.59±0.38) mm. The total difference was statistically significant ( F=281.69, P<0.001). The horizontal corneal diameter measured by IOLMater was (0.49±0.32) mm, which was larger than the horizontal sulcus to sulcus diameter (95% limits of agreement: -0.14, 1.13); the difference between the horizontal corneal diameter measured with OPD-Scan Ⅲ and the horizontal sulcus to sulcus diameter was (0.34±0.32) mm (95% LoA: -0.29, 0.98); and the difference between the horizontal corneal diameter measured with Pentacam and the horizontal sulcus to sulcus diameter was (0.08±0.29) mm (95% LoA: -0.49, 0.66). The difference between the horizontal corneal diameter and the horizontal sulcus to the sulcus was positively correlated with the mean keratometry and negatively correlated with the axial length of the eye (all P<0.01). Conclusion:There are differences between the horizontal corneal diameter measured by the three devices and the horizontal ciliary sulcus diameter, among which Pentacam is the closest, and the differences are correlated with mean keratometry and axial length.

More
广告
  • 浏览55
  • 下载2
中华眼外伤职业眼病杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷