胸部能谱成像模式与常规CT扫描的辐射剂量及图像质量的仿真体模研究

A study on radiation dose and image quality of chest spectral imaging and conventional CT in anthropomorphic phantom

摘要目的 通过仿真胸部体模研究胸部能谱CT不同方案成像模式与常规胸部CT扫描的辐射剂量、图像质量、对比噪声比及主观评分对比,获得最佳能谱扫描参数.方法 对仿真胸部体模分别进行常规胸部CT扫描以及3种不同扫描方式的能谱CT成像.3种能谱模式为宝石能谱CT(GSI) Assist模式(方案A)、管电流平均值时的GSI模式(方案B)及管电流最大值时的GSI模式(方案C).所有扫描方式分别在噪声指数(NI) =9和11时,螺距0.984∶1,依次扫描.记录辐射剂量,同时测量感兴趣区(ROIs)5个不同层面水平的脂肪和肌肉组织的图像噪声值(SD)以评价图像质量.所有扫描序列由两位资深放射医师对肺窗肺纹理及分支5分制主观评分.结果 NI =9和11时,常规CT平扫与方案A、B、C的有效剂量(E)值分别为(8.0、8.5、6.2、10.4)和(5.3、5.1、4.3、6.2) mSv.NI =9时,常规胸部CT平扫与方案A、C的SD值差异有统计学意义(F=4.496,P <0.05);NI=11时,方案A、B、C与常规胸部CT平扫SD值差异有统计学意义(F =8.425,P<0.05);常规胸部CT扫描中,NI分别为9和11时SD值差异有统计学意义(t=-2.570,P<0.05);在相同能谱扫描模式中,NI不同,SD值差异均无统计学意义(P>0.05).NI相同扫描模式不同及扫描模式相同NI不同时,CNR及主观评分差异均无统计学意义(P>0.05).结论 合理的能谱扫描模式与常规扫描的辐射剂量没有明显差异,但能谱扫描模式可以获得较高的图像质量.此外,选择合适的噪声指数在图像质量相仿的同时可以明显降低辐射剂量.综合辐射剂量及图像质量,能谱CT智能模式可以达到辐射剂量及图像质量的双向平衡.

更多

abstractsObjective To obtain the parameters of optimal spectral scanning mode in chest spectral CT imaging,radiation doses,image quality,CNR and subjective scores are compared in spectral and conventional scan modes with anthropomorphic chest phantom.Methods An anthropomorphic chest phantom underwent both conventional scan and spectral scan using three different protocols:GSI assist mode (protocol A);GSI mode with mean tube current (protocol B);GSI mode with maximum tube current (protocol C).All above scans were performed with the noise index (NI) as 9 and 11 respectively.The radiation dose was recorded,and the SD values of adipose tissue and muscle tissue of 5 different sections were measured in regions of interest (ROls) to evaluate the image quality.All scans were subjectively scored in 5 score system by two senior radiologists on lung markings and branches with the lung window technique.Results The effective dose (E) values of conventional CT scan and protocols A,B and C were 8.0,8.5,6.2,10.4 and 5.3,5.1,4.3,6.2 mSv.When NI =9,the difference of SD values had statistically significance between conventional scan and protocols A,C (F =4.496,P < 0.05).When NI =11,there was statistically significant difference of SD value between conventional scan and protocols A,B,C (F =8.425,P < 0.05).In conventional chest scan,the difference of SD values was statistically significant between NI =9 and 11 (t =-2.570,P < 0.05),while other protocols had no statistically significant difference (P > 0.05).The difference of CNR and subjective score had no statistically significance in either the same NI different scanning modes or the same scanning mode with different NIs.Conclusions Appropriate spectral scanning mode had no significant difference with conventional CT scanning in radiation dose,yet higher image quality can be obtained.Furthermore,appropriate NI can reduce the radiation dose obviously while the similar image quality was achieved.Considering the two factors,GSI assist mode can achieve balance between radiation dose and image quality.

More
广告
作者 徐秋贞 [1] 邵海磊 [1] 吕燕 [1] 郭文文 [1] 学术成果认领
作者单位 东南大学医学院 东南大学附属中大医院放射科, 南京,210009 [1]
栏目名称
DOI 10.3760/cma.j.issn.0254-5098.2017.12.015
发布时间 2018-01-23(万方平台首次上网日期,不代表论文的发表时间)
  • 浏览226
  • 下载146
中华放射医学与防护杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷