LabSOCS无源效率刻度软件验证与测量研究

Validation of the Laboratory Sourceless Calibration Software (LabSOCS) on measurement

摘要目的 检验LabSOCS无源效率刻度软件在实际样品测量中的应用效果,对样品自吸收校正和级联辐射符合相加修正进行分析.方法 利用实验室γ谱仪对某公司的两个标准样品进行一系列的测量.采用LabSOCS无源效率刻度软件生成的效率对测量结果进行计算.与证书标准值进行比较,验证分析各核素的计算结果.结果 LabSOCS无源效率刻度软件可以较好地解决γ能谱测量分析中的样品自吸收校正问题,宽能型高纯锗γ能谱仪采用LabSOCS分析的241 Am和109Cd结果与标准值的最大偏差<4%,但对于级联辐射引起的符合相加修正还不能直接解决.结论 LabSOCS无源效率刻度软件通过对样品的成分、密度等信息的精准描述,可以较好地解决样品自吸收校正问题.但当分析带有级联辐射的核素时,仍需对符合相加效应加以考虑.可以利用LabSOCS灵活的模拟方式,增加带有级联核素样品到探测器的距离,完成级联核素的精准测量.

更多

abstractsObjective To validate the effect of the Laboratory Sourceless Calibration Software (LabSOCS) in the actual sample measurement,and to study the effect of sample self-absorption correction and cascade radiation coincidence correction.Methods Two standard samples from the analytics company were measured using the two gamma spectrometers at our laboratory.The result were calculated using the efficiency generated by LabSOCS.Calculated result of each nuclide were analyzed.Results were compared with the standard value provided in the certificate.Results The LabSOCS may solve the problem of sample self-absorption correction in gamma energy spectrum measurement and analysis.The maximum deviation between the values of 241Am and 109Cd analyzed by LabSOCS and the reference value was less than 4% for the broad-energy high-purity germanium gamma-ray spectrometers.However,the coincidence additive correction caused by cascade radiation could not be solved directly.Conclusions The research shows that the LabSOCS can solve the problem of sample self-absorption correction by describing the composition and density of the samples accurately.However,when analyzing the nuclides with cascade radiation,the coincidence addition effects should be considered.The flexible simulation method of the LabSOCS can be used to increase the distance from the cascade nuclide sample to the detector and to complete the accurate measurement of the cascade nuclide.

More
广告
  • 浏览0
  • 下载0
中华放射医学与防护杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new智医 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷