定位像扫描参数对胸部CT影像质量和辐射剂量影响的模体研究

The influence of scout scanning parameters on image quality and radiation dose of chest CT: a phantom study

摘要目的:探讨胸部CT定位像扫描参数(X射线管投照角度和管电压)的选择对图像质量和辐射剂量的影响规律,为临床实践中成像参数的选择提供指导。方法:选择不同扫描参数对成人胸部拟人模体进行定位像采集,X射线管投照角度(0°、90°和180°)、管电压(70、80、100、120和140 kV)和管电流(25 mA)共15种组合。根据所得定位像进行胸部螺旋扫描,扫描条件为Assist kV、Smart mA(设备允许最大范围)、探测器宽度80 mm,螺距0.992∶1,旋转时间0.5 s,扫描长度330 mm, 层厚5 mm,噪声指数(NI)10,迭代指数(ASiR-V)前置为30%,后置50%。记录螺旋扫描管电压、4个层面(肺尖、气管分叉、乳腺水平和横膈顶部)管电流量和容积CT剂量指数(CTDI vol)。用热释光剂量计(TLD)测量每次扫描时乳腺器官剂量。所得图像在气管分叉和横膈顶部层面选取感兴趣区(ROI),计算对比噪声比(CNR)。 结果:X射线管投照角度0°时,螺旋扫描管电压自动选择80 kV,定位像管电压改变对4个层面管电流量影响较小,变化范围0~2%(5/230)。X射线管投照角度90°和180°时,螺旋扫描管电压自动选择100 kV,定位像管电压改变对气管分叉层面管电流量影响较大,变化范围14%(29/210)~44%(93/210)。X射线管投照不同角度其CTDI vol差异有统计学意义( P<0.017);乳腺器官剂量、气管分叉层面CNR和横膈顶部层面CNR差异有统计学意义( F=13.027、24.727和10.630, P<0.05)。根据定位像管电压分组,CTDI vol、乳腺器官剂量和两个层面的CNR差异均无统计学意义( P>0.05)。 结论:胸部CT扫描中,定位像参数中X射线管投照角度较管电压对图像质量和辐射剂量影响更显著。

更多

abstractsObjective:To explore the effects of scout scanning parameters (projection angle and tube potential) on image quality and radiation dose in chest CT and to provide guidance for the selection of scanning parameters in clinical practice.Methods:Different parameter combinations were selected to acquire the scout view of the anthropomorphic adult chest phantom. Fifteen scout views were obtained under parameter combinations of the projection angle (0/90/180 degrees), tube potential (70/80/100/120/140 kV) and tube current (25 mA). And then, chest spiral scans were performed according to the scout views. The other parameters were Assist kV, smart mA (maximum range), 80 mm detector width, 0.992∶1 pitch, 0.5 s rotation time, 330 mm scanning length, 5 mm thickness, NI (noise index) 10. The radiation dose modulation index (ASiR-V) was 30%, and the iterative level was 50%. Volume CT dose index (CTDI vol) and tube current at 4 body levels (pulmonary apex, trachea bifurcation, mammary level and diaphragmatic dome) were recorded. The thermoluminescent dosimeters (TLDs) were used to measure the organ dose of the breast glandular. Regions of interest (ROIs)were selected at the trachea bifurcation and diaphragmatic dome to calculate the contrast noise ratio (CNR). Results:The tube potential of the spiral scanning was automatically selected as 80 kV when the projection angle was 0 degree. The changing of the tube potential for the scout had little influence on the tube current at the four body levels, and the variation range was 0-2% (5/230). The tube potential in the spiral scanning was automatically selected as 100 kV when the projection angles were 90 and 180 degrees. The changing of the tube potential for the scout had a great influence on the tube current at the slice of trachea bifurcation and the variation range was from 14%(29/210)to 44%(93/210). According to the angle projection, there were statistical differences in CTDI vol( P< 0.017), organ dose of the breast glandular, and CNR at the slices of trachea bifurcation and at diaphragmatic dome ( F=13.027, 24.727, 10.630, P< 0.05). According to the tube potential in the scout scanning, there were no significant differences in CTDI vol, organ dose of the breast glandular and CNR at both levels ( P> 0.05). Conclusions:In chest CT, the influence of the projection angle in scout scanning on the image quality and radiation dose was more significant than that of the tube potential.

More
广告
栏目名称
DOI 10.3760/cma.j.issn.0254-5098.2021.03.011
发布时间 2021-03-25(万方平台首次上网日期,不代表论文的发表时间)
  • 浏览0
  • 下载0
中华放射医学与防护杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new智医 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷