比较头颈部肿瘤IMRT中XGS-1O与OBI系统的摆位误差

Analysis of the setup errors of a stereoscopic two-dimensional kilo-voltage XGS-10 system for head-and-neck region intensity-modulated radiotherapy

摘要目的 比较采用千伏级XGS-10与OBI系统在头颈部肿瘤IMRT中的摆位误差和图像采集、配准所需时间差异.方法 30例头颈部患者IMRT前分别采用OBI系统获取CBCT图像及XGS-10系统获取X线图像,将2套图像与计划图像及其靶中心匹配,分别得出靶中心左右、上下、前后方向摆位误差数据.采用Pearson法相关分析2套系统分别在3个方向上的相关性及一致性的可信区间.结果 左右、上下、前后方向摆位误差XGS-10系统的分别为(-1.03 ±2.15)、(0.86±2.59)、(0.42±1.66) mm,OBI系统的分别为(0.00±1.68)、(1.53±2.12)、(0.10±1.54) mm,二者差值分别为(-1.03±1.24)、(-0.68±1.78)、(0.32±1.61) mm.二者在左右、上下、前后方向的相关系数分别为0.817、0.731、0.495,95%CI值分别为(-1.47~-0.59)、(-1.32~0.04)、(-0.26 ~0.90)mm.XGS-10获取图像、配准平均时间分别为10s和<15s,OBI的需3 min和8 min.结论 XGS-10系统与OBI系统均可提高摆位精度,但XGS-10系统可缩短摆位及摆位误差纠正时间.

更多

abstractsObjective To analyze the discrepancies between position adjustments obtained with the stereoscopic 2DKV XGS-10 system and the Varian OBI system for head-and-neck region IMRT treatments,and to compare for image acquisition and registration time.Methods CBCT images were obtained with OBI system and 2DKV images were acquired by XGS-10 system for 30 head-and-neck patients prior to Varian21EX IMRT treatment.The images were registered with planning image for localization,and position adjustments were given in LR,SI and AP directions,then the discrepancies between them were analyzed.On the comparison of the two different systems,the Pearson coefficient was used to analyzed the correlation and 95% CI analysis to discern the consistence.Results Analysis of images acquired for the 30 patients yielded the following results:position adjustments with XGS-10 system were (-1.03 ± 2.15) mm,(0.86 ± 2.59) mm,(0.42 ± 1.66) mm in LR,SI and AP directions,whereas (0.00 ± 1.68) mm,(1.53 ± 2.12) mm,(0.10 ± 1.54) mm with CBCT in LR,SI and AP directions.The discrepancies were (-1.03 ± 1.24) mm,(-0.68 ± 1.78) mm and (0.32±1.61) mm in LR,SI and AP directions.The correlation coefficients between them were 0.817,0.731 and 0.495 in LR,SI and AP directions.95% CI were (-1.47--0.59),(-1.32-0.04),(-0.26-0.90) mm.The average image acquisition and registration time were 10 s and < 15 s in XGS-10 system,with 3 min and 8 min in OBI system.Conclusions Both of XGS-10 system and OBI system could be used to improve patient position accuracy,but XGS-10 system could cut down the total time.

More
广告
栏目名称
DOI 10.3760/cma.j.issn.1004-4221.2014.03.020
发布时间 2014-05-29(万方平台首次上网日期,不代表论文的发表时间)
  • 浏览256
  • 下载60
中华放射肿瘤学杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷