两种调强治疗计划系统的Axesse加速器光子束模型验收测试

Acceptance testing of photon beam model in Axesse accelerator in two intensity-modulated treatment planning systems

摘要目的 研究Monaco、OncentraTPS的Axesse加速器光子束模型的验收测试,以期确定两者同时验收的完整方法.方法 验收前考虑两TPS相矛盾的参数,将加速器对应参数调至两TPS的交集部分;验收分为点、面、MLC特性模拟测试,MLC特性测试包括凹凸槽、穿射、“offset”、“Gain”值等.结果 绝大部分点剂量误差在4%以内,只有部分半影区及射野外测量点偏差>4%.面剂量及所有调强病例在3mm/3%时的γ通过率除3个测试例外均>90%.Oncentra只能模拟MLC的Offset值为正的情况,Monaco对Offset值为正、负时的情况均能模拟.结论 结果显示该方法能准确的验证两TPS各项指标,能够实现对Axesse加速器Monaco、Oncentra TPS的准确验收;Monaco、Oncentra模型均满足临床计算要求.

更多

abstractsObjective To investigate the acceptance testing of the photon beam model in Axesse accelerator in Monaco and Oncentra treatment planning systems (TPS),in order to establish an integral method for the acceptance of both systems at the same time.Methods The conflicting parameters of the two TPSs were determined before acceptance,and the corresponding parameters in the accelerator were adjusted to the overlapping values.The acceptance test was performed from the aspects of point dose,surface dose,and characteristic simulation of multi-leaf collimator (MLC).The test case for point dose was designed based on TRS430,TG53,and practical situations.The test case for surface dose included conventional radiation field and intensity-modulated radiotherapy (IMRT) cases.The test of the characteristics of MLC included concave-convex groove,leakage,and values of" offset" and" Gain".Results The error of most point doses were within 4%,and only some measuring points in the penumbra region and outside the radiation field had an error of>4%.The γ pass rate of the surface dose tests and IMRT cases was>90% at a value of 3 mm/3%except three test cases.Oncentra could only simulate the situation with a positive offset value of MLC,but Monaco could simulate the situation with both positive and negative offset values of MLC.Conclusions The results show that this method can accurately validate all the parameters of the two systems and realize the accurate acceptance of Monaco and Oncentra systems with Axesse accelerator.Both Monaco and Oncentra models can meet the needs for clinical calculation.

More
广告
  • 浏览211
  • 下载116
中华放射肿瘤学杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷