OBI图像引导在鼻咽癌IMRT中的应用

Application of OBI system in intensity-modulated radiotherapy for nasopharyngeal carcinoma

摘要目的 探索IGRT在保障鼻咽癌IMRT精确性中的应用.方法 应用直线加速器OBI系统对10例鼻咽癌患者实施IMRT,观察并记录每次摆位后、体位调整后、治疗后的IGRT,分析其影像与CT模拟定位影像之间的摆位差异.结果 摆位后左右、头脚、前后方向误差分别为(0.22±1.00)、(-0.37±1.28)、(0.04±1.36) mm;体位调整后对应为(0.29±0.76)、(-0.04±0.78)、(-0.01±0.92) mm;治疗后对应为(0.20±0.78)、(0.16±0.80)、(0.05±0.92) mm.摆位后左右、头脚、前后方向误差≤1 mm概率分别为81.0%、77.6%、88.2%;体位调整后对应为92.5%、96.4%、96.4%;治疗后对应为91.7%、94.9%、96.8%.结论 鼻咽癌患者实施分次IMRT过程中进行OBI系统下的精准IGRT技术非常重要,必要时需参照摆位后的IGRT影像调整患者的放疗体位,修正摆位误差,有效提高分次IMRT的精确性.

更多

abstractsObjective To examine the application of On-Board Imaging (OBI) system-based image-guided radiotherapy (IGRT) in the improvement of the precision of intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma.Methods Ten patients with nasopharyngeal carcinoma were treated with IMRT using the OBI system. The IGRT images after positioning, position adjustment, and treatment were observed and recorded to investigate the image difference between CT simulation and IGRT. Results The deviations in the x (lateral), y (cranial-caudal), and z (ventral-dorsal) directions between CT simulation and IGRT images were 0.22±1.00 mm,-0.37±1.28 mm, and 0.04±1.36 mm, respectively, after positioning, 0.29±0.76 mm,-0.04±0.78 mm, and -0.01±0.92 mm, respectively, after position adjustment, and 0.20±0.78 mm, 0.16±0.80 mm, and 0.05±0.92 mm, respectively, after treatment. The probabilities of a ≤1 mm deviation in the x, y, and z directions were 81.0%, 77.6%, and 88.2%, respectively, after positioning, 92.5%, 96.4%, and 96.4%, respectively, after position adjustment, and 91.7%, 94.9%, and 96.8%, respectively, after treatment. Conclusions The application of OBI system-based IGRT is very important in the improvement of the precision of fractionated IMRT for patients with nasopharyngeal carcinoma. The position of the patient should be adjusted based on the IGRT image after positioning in order to correct set-up error and effectively increase the precision of fractionated IMRT.

More
广告
  • 浏览289
  • 下载124
中华放射肿瘤学杂志

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷