天平室温度和湿度的变化对测尘滤膜称量结果的影响

Effect of humidity and temperature on filter and gravimetric measurement of ambient particulate matter in a balance room

摘要目的 研究称量环境温度和湿度的变化对采样前后滤膜称量结果的影响.方法 在(18±1)℃和(28±1)℃2个温度下,调节天平室的湿度,于不同气象条件下对采集煤尘前后的测尘滤膜进行称量,计算采样后承载滤膜的增量(相同称量气象条件下的承载滤膜质量减去相对应的空白滤膜的质量),对结果进行比较分析.结果 空白滤膜和承载滤膜在(18±1)℃和(28±1)℃的称量结果波动范围分别为0.10~0.13 mg和0.06~0.09 mg,温度和湿度均变化时,称量结果波动范围为0.12~0.16 mg;各组承载滤膜增量均值在(18±1)℃和(28±1)℃的波动范围分别为0.07~0.10 mg和0.04~0.08 mg,温度和湿度均变化时,波动范围为0.09~0.14 mg.同一温度组内,空白滤膜组和承载滤膜组的称量均值随着湿度升高呈上升趋势,而各承载滤膜的增量与湿度不相关;在相同湿度下,(18±1)℃时承载滤膜增量高于(28±1)℃,差异有统计学意义(P<0.01);除总尘高增量组外,其余各组称量均值在(18±1)℃随湿度升高的上升趋势比(28±1)℃更明显,差异有统计学意义(P<0.01).当湿度相同时,各承载滤膜组(18±1)℃的增量均值高于(28±1)℃,差异有统计学意义(P<0.01).结论 实验室温度和湿度的变化导致测尘滤膜称量结果的波动,湿度对采样前后滤膜增量影响不明显,但低温时滤膜增量高于高温时,会影响粉尘浓度测定结果.

更多

abstractsObjective To assess the effects of the alteration of humidity and(or)temperature on weight of filters without and with ambient particulate matter in a balance room.Methods The mass of blank dust sampling filters were weighed under(18±1)℃ and(28±1)℃ respectively,with the humidity varying from 35% relative humidity(RH) to 100% RH in a balance room.Then the blank filters were divided into two groups and were used to sample total dust and respirable dust.After sampling,the loaded filters were reweighed under above conditions and the mass difference before and after the sampling were compared and analyzed.Results The vibration of the average mass of filters varied from 0.10 to 0.13 mg and from 0.06 to 0.09 mg under the temperatures of(18±1)℃ and(28±1)℃ respectively;When both the temperature and humidity changed,it varied from 0.12 to 0.16 mg.The deviation of average mass difference ranged from 0.07 to 0.10 mg and from 0.04 to 0.08 mg under the two temperatures mentioned above:When both the temperature and humidity changed,it varied from 0.09 to 0.14 mg.The average mass of blank filters and loaded filters were all positively correlated with the change of humidity (P<0.01).No effects of humidity on the average mass difference of the loaded filters were observed.The average mass differences of loaded filters and blank filters under(18±1)℃ were significantly higher than that under(28±1)℃(P<0.01) when humidity was not changed.Conclusion The alteration of humidity and (or) temperature in a balance room attributes to the deviation of the measurement of the mass of filters and thus affects the gravimetric measurements of ambient particulate matter.

More
广告
  • 浏览603
  • 下载8
中华劳动卫生职业病杂志

中华劳动卫生职业病杂志

2008年26卷4期

198-202页

MEDLINEISTICPKUCSCDCA

加载中!

相似文献

  • 中文期刊
  • 外文期刊
  • 学位论文
  • 会议论文

加载中!

加载中!

加载中!

加载中!

扩展文献

法律状态公告日 法律状态 法律状态信息

特别提示:本网站仅提供医学学术资源服务,不销售任何药品和器械,有关药品和器械的销售信息,请查阅其他网站。

  • 客服热线:4000-115-888 转3 (周一至周五:8:00至17:00)

  • |
  • 客服邮箱:yiyao@wanfangdata.com.cn

  • 违法和不良信息举报电话:4000-115-888,举报邮箱:problem@wanfangdata.com.cn,举报专区

官方微信
万方医学小程序
new医文AI 翻译 充值 订阅 收藏 移动端

官方微信

万方医学小程序

使用
帮助
Alternate Text
调查问卷